IOLMaster 700 product photo front 5XL
000000-2118-007-000127

IOLMaster 700

  Technical specifications of IOLMaster 700 by ZEISS
Measurement range Axial length 14 – 38 mm
Corneal radii 5 – 10 mm
Anterior chamber depth 1,5 – 6,5 mm
White-to-white 8 –16 mm
Screen scale Axial length 0,01 mm
Corneal radii 0,01mm
Anterior chamber depth 0,01 mm
Lens thickness 0,01 mm
Central corneal thickness 1 μm
White-to-white 0,1mm
SD of repeatability Axial length 8 μm
Corneal radii 0,09 D
Cylinder > 2,0 D each 1,7°
Anterior chamber depth 11 μm
Lens thickness 12 μm
Central corneal thickness 2 μm
IOL calculation formulas SRK® / T, Holladay 2, H offer Q, Haigis Suite (includes Haigis,
Haigis-L for eyes after LASIK surgery
PRK / LASEK for myopia / hyperopia and Haigis-T for calculation
of toric IOL power)
Interfaces ZEISS FORUM® Ophthalmology Data Management System
ZEISS computer-assisted cataract surgery system CALLISTO eye (via USB & FORUM)
Data interface for electronic medical record (EMR) / patient management systems (PMS)
Data export to USB storage media
Ethernet port for network connection and network printer
Line voltage 100 – 240 V ± 10 % (auto detection)
Line frequency 50 – 60 Hz
Consumption max. 150 VA
Laser class 1